Z-0925-2021: Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healt

Recall record Z-0925-2021 for Hill-Rom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0925-2021
Company in FDA record
Hill-Rom, Inc.
FDA classification date
Jan 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming detached from the rail system, potentially resulting in the fall of both the lift motor and the patient

Source facts

Action
Hill Rom issued "Urgent Medical Device Correction" notice on 18 December 2020 to Distributors and End Users. The letter states: Hillrom will perform this field action in two phases. In the interim, you may continue to utilize your medical device. Hillrom will be replacing the Q-link on Multiralls to Q-link II to improve ease of use and mitigate potential risk to patient or caregiver. Phase 1: Users inspect their Multirall installations, fill out the response forms and return it to Hillrom or Hillrom distributor. The serial numbers are stated on the product label attached to the lift. Kindly read carefully and take the 2 actions within one month: Action 1: Please inspect each Multirall installation in your facility and identify which category (A, B or C) it belongs to: Category A Multirall installations are used in combination with any of the room to room accessories, Universal SlingBar 350 R2R or Universal SlingBar 450 R2R or room to room rail carriage D45 with Double Hook . Category B Multirall installations are combined with an extension belt. Category C Multirall installations are not combined with any of the accessories in A or B Action 2: Please fill out the attached response form and return it to Hillrom (MOD1322@hillrom.com) within one month. This will provide us with the necessary information to plan phase 2 of the field action. Phase 2: Hillrom or an official Hillrom distributor will contact you to plan replacement. Once you have identified units affected by this field corrective action and you have returned the response form, you will be contacted by Hillrom or an official Hillrom distributor, to schedule the replacement of the Q-link 1, R2R accessories and Extension belts if present in the installations. Please transfer this notice to other organizations as appropriate. Please maintain awareness of this notice and additional resource for an appropriate period to ensure effectiveness. If you have any questions regarding this safety notice
Initiated
2020-12-18
Root cause
Device Design
Terminated
2023-09-26
Distribution
Worldwide distribution: US (Nationwide) and countries of::AT, AU, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, GR, IE, IL, IS, IT, JP, LV, NL, NO, NZ, SE & SK.
Recall number
Z-0925-2021
Recall status
Terminated
CDRH record ID
185135
FDA classified
2021-01-29
Recall event number
87053

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗