Open original FDA record ↗Recall record
- FDA identifier
- Z-0926-2014
- Company in FDA record
- ArjoHuntleigh Inc.
- FDA classification date
- Feb 4, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The lock pin can become stuck in the "in" position during the manual rotation, resulting in the inability to prone a patient
Source facts
- Action
- ArjoHuntleigh Inc. sent a Technical Bulletin package to all affected service centers and an Urgent Field Safety Notice package to all affected customers. For rental units, a training plan was established for employees. For sold units, the notice contain a revised user manual and label explaining what to do if the lock pin becomes stuck in the "in" position. Customers that purchased the affected product were required to complete and return a Customer Response form to ArjoHuntleigh to confirm receipt and understanding of the field safety notice, and that the required actions were completed. For questions call 630-785-4885.
- Initiated
- 2013-09-11
- Root cause
- Device Design
- Terminated
- 2015-01-08
- Distribution
- US Nationwide and Middle East
- Recall number
- Z-0926-2014
- Recall status
- Terminated
- CDRH record ID
- 125084
- FDA classified
- 2014-02-04
- Recall event number
- 67248