Skip to content

Public FDA record

Z-0926-2023: SHIP KIT, LIFESPARC, CONTROLLER AND ACCESSORIES - 5900-0000; SHIP KIT, LIFESPARC, CONTROLLER ONLY - 5900

Recall record Z-0926-2023 for Cardiac Assist, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0926-2023
Company in FDA record
Cardiac Assist, Inc
FDA classification date
Jan 19, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022. During the Critical Failure, the software freezes or crashes and the screen does not display data.

Source facts

Action
On December 5, 2022, the firm notified customers through Urgent Medical Device Communication letters that a software update has been developed to address the issue of Critical Failure. The only action required by the customer will be to make their equipment available for the third-party representative from MERA to perform the update and sign off on the form once the upgrade is completed. This software update is the follow-up to the earlier recall notification on or around July 21, 2022, when customers were instructed to use a new revision of the Operations Manual. For questions regarding this Medical Device Removal, please contact your LivaNova representative, or LivaNova Customer Quality at LivaNova.FSCA@livanova.com.
Initiated
2022-12-05
Root cause
Software design
Terminated
2023-08-08
Distribution
US Nationwide distribution.
Recall number
Z-0926-2023
Recall status
Terminated
CDRH record ID
197572
FDA classified
2023-01-19
Recall event number
91362

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗