Reason stated in the source
If a plan contains a sequence of beams, including at least 1 ARC treatment beam and if during the setup stage of a beam following an ARC beam in the delivery sequence a termination occurs, and the user selects the immediate resumption feature, syngo RT Therapist Express systems will display and record a second delivery of a prior beam and not the delivery of the current beam in certain circumstanc
Source facts
- Action
- An Urgent Medical Device Customer Safety Letter dated October 23, 2008 was issued to consignees via certified mail. The letter identified the issue and preventive measures for consignees. Monitoring will be done via return cards. Follow-up to be conducted via telephone.
- Initiated
- 2008-10-23
- Root cause
- Software design
- Terminated
- 2010-12-09
- Distribution
- Worldwide Distribution including USA and countries of Netherlands, Germany, Poland, Norway, Australia, Hungary, Belgium, New Zealand, France, Malaysia, Spain, and Canada.
- Recall number
- Z-0927-2009
- Recall status
- Terminated
- CDRH record ID
- 75567
- FDA classified
- 2009-01-29
- Recall event number
- 50515