Skip to content

Public FDA record

Z-0927-2023: DeRoyal ORGAN RECOVERY PACK PGYBK, a) REF 89-9004.09 and b) 89-9004.10

Recall record Z-0927-2023 for DeRoyal Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0927-2023
Company in FDA record
DeRoyal Industries Inc
FDA classification date
Jan 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

DeRoyal is recalling certain lots of surgical procedure pack products that contain a Medtronic Non-Absorbable Suture. The recall is due to Medtronic determining a manufacturing error in the packaging of the non-absorbable sutures that may cause a gap or wrinkle in the breather seal resulting in a sterile barrier breach.

Source facts

Action
DeRoyal sent an urgent recall notice to its consignees on 12/12/2022 by email and they followed with a letter sent overnight to the end user. The notice explained the problem and the hazard, and requested the following actions be taken: "1) Using the attached Affected Products Listing, identify affected surgical pack numbers in your inventory and place in quarantine to prevent further use. Please return the packs that you have in inventory to us. You will be given credit or replacement for the returned product. 2) Complete the NOTICE OF RETURN FORM, indicating all affected products found in your inventory. Please return the completed form to DeRoyal via fax to 865-362-3716 or recalls@deroyal.com no later than January 15, 2023. 3) If you have questions regarding credit for the recall, please contact Jennifer Marsh at jmarsh@deroyal.com or by phone 865-362-4203 or email us at recalls@deroyal.com. 4) PLEASE COMPLETE AND RETURN THIS FORM EVEN IF YOU NO LONGER HAVE ANY INVENTORY. 5) If you have further distributed these products, it is required by federal law that your customers be notified of this recall. You may either do this directly by forwarding our recall notification and notify us that you have done so, or you may provide us with your customer listing and we will contact them."
Initiated
2022-12-12
Root cause
Packaging change control
Terminated
2026-05-29
Distribution
US Nationwide distribution in the states of IL, MO.
Recall number
Z-0927-2023
Recall status
Terminated
CDRH record ID
197494
FDA classified
2023-01-09
Recall event number
91331

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗