Z-0927-2024: Cryo Module Accessories Domestic CMA-OUS, Product Catalog Number A001053

Recall record Z-0927-2024 for AtriCure, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0927-2024
Company in FDA record
AtriCure, Inc.
FDA classification date
Jan 31, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.

Source facts

Action
An URGENT: MEDICAL DEVICE RECALL notification letter dated 11/28/23 was sent to customers. Actions to be taken by the Customer/User: " If the Tank Hose assembly is part of the lots listed below, do not use the tank hose assembly. " If the lot number is listed below, work with your AtriCure sales rep to replace the device with a new Tank Hose assembly and return the recalled device. " Return the attached Medical Device Recall Return Response Acknowledgement and Receipt Form. Type of Action by the Company: AtriCure is arranging for the removal and replacement of the affected tank hose assemblies. If you have impacted product, AtriCure will provide a replacement product at no charge to you. If you have any questions, please contact Rob Cantu, Vice President of Quality at (1-513-755-4100) from 9-5pm ET on Mondays - Fridays. You may also contact customer service at (1-866-349-2342) any time of day, your message will be forwarded to Quality Assurance for prompt review.
Initiated
2023-11-28
Root cause
Other
Distribution
US Nationwide. Brazil, Canada, Japan, Taiwan
Recall number
Z-0927-2024
Recall status
Open, Classified
CDRH record ID
205085
FDA classified
2024-01-31
Recall event number
93672

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗