A significant level of 25G spinal needle hub leakage complaints was observed. Subsequent investigation found the leakage to be attributable to improper formation of the spinal needle hub, specifically an undersized taper and the prescence of a slight sink in the taper surface.
Source facts
Action
The firm sent out notification letters on 3/16/2005.
Initiated
2005-03-16
Root cause
Other
Terminated
2005-11-07
Distribution
The firm uses distributors and ships the product directly to over 4000 hospitals and clinics nationwide.
Recall number
Z-0928-05
Recall status
Terminated
CDRH record ID
38306
FDA classified
2005-06-25
Recall event number
31657