Z-0928-06: Model Number: GNRADIOLUCENTGUIDE (Gliding Nail Radiolucent Guide), instrument set that contains one unit

Recall record Z-0928-06 for Plus Orthopedics USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0928-06
Company in FDA record
Plus Orthopedics USA
FDA classification date
Jun 2, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

PLUS USA has been informed by PLUS AG that the accuracy of the instrument set may be affected in some cases. It is possible that in rare cases where patients have sclerotic (excessively hardened or firm) bone, that components may deform under application of force as the surgeon attempts to penetrate this bone with the instruments.

Source facts

Action
There were a total of five (5) radiolucent guide instrument sets affected by the recall (imported to the U.S. between 1/23/2006 and 1/30/2006). The firm telephoned agents beginning February 22, 2006, requesting return of the devices. Returned Material Authorizations have been issued (March, 2006) for consigned inventory in the field with instructions given to remove and return affected units immediately. All returns to PLUS Orthopedics USA have been appropriately quarantined and segregated to prevent unintended use or shipment.
Initiated
2006-02-22
Root cause
Other
Terminated
2007-12-13
Distribution
There were a total of five (5) radiolucent guide instrument sets affected by the recall (imported to the U.S. between 1/23/2006 and 1/30/2006). The 5 sets were distributed in the following states: 1) California, 2) Idaho, and 3) New York.
Recall number
Z-0928-06
Recall status
Terminated
CDRH record ID
45198
FDA classified
2006-06-02
Recall event number
35054

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗