Z-0928-2020: BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation

Recall record Z-0928-2020 for Biomerieux Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0928-2020
Company in FDA record
Biomerieux Inc
FDA record creation date
Mar 2, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.

Source facts

Action
The firm initiated their field action by an email to their subsidiaries on 10/22/2019. The subsidiaries then notified their customers by 2-day express via Fed Ex. In the US customer letter, customers are instructed to check their software version. If the VIDAS 3 software version is between 1.2.0 and 1.3.1, customers are instructed to check the archiving mode. If the archiving mode is Automatic When Results are Reported then customers are instructed how to configure the workaround, and to contact bioMerieux if the customer has the incorrect software configuration. bioMerieux then instructs impacted customers to review and interpret results in the context of the overall clinical procedure. Once this is completed to complete the acknowledgment form in Attachment A and return it the local bioMerieux representative and confirm receipt of this notice.
Initiated
2019-10-22
Root cause
Software design
Terminated
2024-08-06
Distribution
US
Recall number
Z-0928-2020
Recall status
Terminated
CDRH record ID
178383
FDA record created
2020-03-02
Recall event number
84464

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗