Z-0928-2022: Atellica CH Calcium_2 (CA_2), Siemens Material Number 11097644

Recall record Z-0928-2022 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0928-2022
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Apr 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

Source facts

Action
Urgent Medical Device Correction notification letters dated 2/9/22 were sent to customers. Actions to be Taken by the Customer Please review this letter with your Medical Director. Perform the instructions provided in the Additional Information section below. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received these products. Additional Information If your laboratory has multiple Atellica CH 930 Analyzers, separate the assays as follows: -LITH_2 from CA_2 and TBil_2 -A1c_E/A1c_H from CA_2 - CA_2 from UN_c (urine only); UN_c serum and plasma are not impacted and can be run on the same analyzer as CA_2 -Fruc from AAT, IgA_2, IgM_2 and Theo If you choose not to separate the assays as indicated above, batch testing of A1c_E, Fruc, LITH_2, and UN_c (urine only) may be considered. Note: An RPC2 wash mitigation must be initiated after completion of A1c_E, Fruc, LITH_2, and prior to UN_c (urine only). Any of the following will initiate the RPC2 wash: -After the Atellica CH 930 Analyzer has been in standby for 12 minutes. - After completion of any Open Channel assay. - Restarting the Atellica CH 930 Analyzer. Refer to the Atellica Solution Online Help for instructions on system restart. If you have any questions, contact your local Siemens Healthineers technical support representative.
Initiated
2022-02-09
Root cause
Device Design
Distribution
Global Distribution. US Nationwide.
Recall number
Z-0928-2022
Recall status
Open, Classified
CDRH record ID
192329
FDA record created
2022-04-14
Recall event number
89740

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗