Complaint investigation concluded that although product quality
requirements stated in the instructions for use were met at lot release, the high background in the ELISA plate, the
strong IgM conjugate, and the low reference control led to an increased risk of equivocal and/or false positive sample
test results for the two FANA200 kit lots SS 1500 and SS 2009.
Source facts
Action
Euro Diagnostica sent a Customer notifications letter to affected customers via email on 10/15/2015. The letter identified the affected product problem and actions to be taken. For question contact Euro Diagnostica .
Initiated
2015-10-15
Root cause
Nonconforming Material/Component
Terminated
2019-03-29
Distribution
Worldwide Distribution and US. Nationwide
Recall number
Z-0929-2018
Recall status
Terminated
CDRH record ID
161552
FDA record created
2018-03-08
Recall event number
79183