Open original FDA record ↗Recall record
- FDA identifier
- Z-0930-2009
- Company in FDA record
- Invacare Corporation
- FDA classification date
- Jan 27, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Invacare has become aware that on some chairs, the stability lock feature may not be engaging properly or consistently. If the stability lock feature does not engage properly, the wheelchair may veer to one side or it may rock forward onto its front riggings.
Source facts
- Action
- On 11/27/2007 the firm sent Urgent Field Correction Needed recall notification letters and response forms to their customers informing them of the problem. They instructed their customers to immediately inspect and adjust the Stability Lock feature, distribute the enclosed update to the Owner's Manual to all end-users, and return the enclosed response form notifying them that the inspection and/or adjustments have been completed.
- Initiated
- 2007-11-27
- Root cause
- Pending
- Terminated
- 2013-08-01
- Distribution
- Worldwide Distribution --- including USA and countries of Canada, Australia, Germany, New Zealand and the United Kingdom.
- Recall number
- Z-0930-2009
- Recall status
- Terminated
- CDRH record ID
- 74891
- FDA classified
- 2009-01-27
- Recall event number
- 50169