Z-0930-2014: ADVIA Centaur Systems Phenytoin Calibrator N ( PHTN) 6-pack SMN 10314977 For in vitro diagnostic use in

Recall record Z-0930-2014 for Siemens Healthcare Diagnostics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0930-2014
Company in FDA record
Siemens Healthcare Diagnostics, Inc
FDA classification date
Feb 4, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ADVIA Centaur Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots.

Source facts

Action
Siemens Healthcare Diagnostics Customer Notification notice was sent to affected Siemens Healthcare Diagnostic customers in the United States on December 18, 2013.. The notifications inform customers that Siemens Healthcare Diagnostics received customer feedback regarding the performance of the ADVIA Centara Systems Phenytoin assay, indicating that the assay was recovering higher relative to the performance of other assays in proficiency studies. Siemens internal investigation confirmed the positive bias for Calibrator N lot CN 57 in the ADVIA Centaur Systems Phenytoin assay. This lot has expired and is no longer in use. Expected performance has been restored starting with Calibrator N lot CN58. Siemens is notifying customers that have received Calibrator N lots CN58 and higher that there will be a change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots. Siemens is providing data to inform customers of the expected difference between Calibrator lots CN58 and higher versus prior Calibrator N lots. Customers were instructed to retain the letter with their laboratory records, and forward to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Customers Care Center or their local Siemens technical support representative. For questions regarding this recall call 508-668-5000.
Initiated
2013-12-18
Root cause
Under Investigation by firm
Terminated
2016-05-18
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia,Brazil,Canada, Columbia, China, Ecuador, Egypt, India, Japan, Korea, Mexico, Saudi Arabia,Singapore, Thailand, and Uruguay.
Recall number
Z-0930-2014
Recall status
Terminated
CDRH record ID
124759
FDA classified
2014-02-04
Recall event number
67021

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗