IBA identified incorrect screw holes depth in graphite block 8 with respect to the degrader
Source facts
Action
On July 17, 2017 An Urgent Field Safety Notice was issued to customers with the following user action:
In case a patient treatment plan requires the use of the block 8, it is recommended to perform a full range verification during the patient treatment plan QA check. After the replacement of the graphite block 8, it is recommended to verify there is no effect on the Bragg peak.
Initiated
2017-10-17
Root cause
Component design/selection
Terminated
2019-03-29
Distribution
WA
Recall number
Z-0930-2018
Recall status
Terminated
CDRH record ID
161689
FDA record created
2018-03-08
Recall event number
79227