Open original FDA record ↗Recall record
- FDA identifier
- Z-0931-2013
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Mar 10, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
Source facts
- Action
- Elekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated December 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The IFSN informs the user in the TPS, to give the different volumes different structure names. After importing the patient data set from the TPS, make sure that the structure contours are correct. Users are requested to return the signed acknowledgement. Contact your local Elekta office for questions regarding this notice.
- Initiated
- 2012-12-19
- Root cause
- Software design
- Terminated
- 2021-07-14
- Distribution
- Nationwide Distribution-DC, Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, FL, GA, IL, IA, KY, MD, MA, MI, MN, MS, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, and WI.
- Recall number
- Z-0931-2013
- Recall status
- Terminated
- CDRH record ID
- 115474
- FDA classified
- 2013-03-10
- Recall event number
- 64165