Range for the same energy could be different depending if the layer is
the first layer of the irradiation plan or not.
Source facts
Action
IBA sent an Urgent Field Safety Notice dated October 19, 2017 , issued to customers stating the following: IBA recommends performing a check to detect the consequence of this potential power supply regulation issue on a regular basis (IBA recommends Monthly), see Annex 1 for details.
IBA is developing a corrective action in order to reduce the hysteresis
effect and detect magnet current overshoot.
This solution will be deployed to all systems using JEMA analog power
supply. This solution will be validated and deployed by April 2018. For further questions, please call (215) 972-7985
Initiated
2017-10-19
Root cause
Device Design
Terminated
2019-04-02
Distribution
US Distribution to the state of :FL
Recall number
Z-0931-2018
Recall status
Terminated
CDRH record ID
161705
FDA record created
2018-03-08
Recall event number
79228