Open original FDA record ↗Recall record
- FDA identifier
- Z-0933-05
- Company in FDA record
- Masimo Corporation
- FDA classification date
- Jun 25, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Speaker component failure resulting in ceasation of an audible alarm even though the condition continues that caused the alarm.
Source facts
- Action
- Recall was by letter to all customers who received units made before 12/11/01. They are to return for repair or have them fixed by an Authorized Masimo Service Representative.
- Initiated
- 2005-06-10
- Root cause
- Other
- Terminated
- 2008-06-17
- Distribution
- Nationwide and Korea, Hong Kong, Canada, New Zealand, Thailand, Argentina, Portugal, Malaysia, Ireland, Australia, japan, France, El Salvador, South Africa, UAE-Saudi Arabia, Germany, Mexico, Unitied Kingdom, Singapore, Germany, Kuwait, Chile, Spain, Iceland, Italy, Poland, Sweden, Norway, Venezuela, Lithuania, Czech Republic, Belgium, Neterlands, Austria, Switzerland, Egypt, Greece, France Taiwan.
- Recall number
- Z-0933-05
- Recall status
- Terminated
- CDRH record ID
- 39936
- FDA classified
- 2005-06-25
- Recall event number
- 32384