Failure to shut-off: Surgical equipment, activated by a wireless foot switch with VASER system, may remain powered on when switch is no longer depressed.
Source facts
Action
Consignees were sent an Advisory Notice on 12/28/2007. The notification describes the defect. Sound Surgical will be contacting facilities to arrange a service call to install new cables between the footswitch receiver and the VASER system. Facilities may contact the Technical Service Center at 1-888-677-8765.
Initiated
2007-12-28
Root cause
Device Design
Terminated
2008-04-01
Distribution
Worldwide Distribution including USA states of AL, CA, CT, FL, GA, IN, KY, LA, MI, NJ, NY, OH, OK, PA, TX, and VA. Foreign Distribution to Colombia, Costa Rica and Switzerland.
Recall number
Z-0933-2008
Recall status
Terminated
CDRH record ID
67151
FDA classified
2008-03-13
Recall event number
46293