Skip to content

Public FDA record

Z-0933-2020: Reinforced Introducer Sets Maquet 7.5 Fr., Part Number: 068400040305. It is an accessory to be used for

Recall record Z-0933-2020 for Datascope Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0933-2020
Company in FDA record
Datascope Corporation
FDA record creation date
Feb 2, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for compromised sterility in breached pouches of specific lots of Reinforced Introducer Sets Maquet 7 Fr., 7.5 Fr. and 8 Fr. IABs.

Source facts

Action
On December 11, 2019 the firm distributed Urgent Medical Device Recall Communication letters by FedEx 2 day delivery with signature proof of delivery stating: URGENT MEDICAL DEVICE RECALL REMOVAL MAQUET 7Fr., 7.5Fr. and 8Fr. REINFORCED INTRODUCER SETS PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL USERS OF REINFORCED INTRODUCER SETS FOR MAQUET 7 Fr., 7.5 Fr. AND 8 Fr. INTRA-AORTIC BALLOON (IAB) WITHIN YOUR HOSPITAL OR FACILITY. IF YOU HAVE FURTHER DISTRIBUTED ANY OF THE AFFECTED PRODUCTS, FORWARD THIS INFORMATION TO THE RECIPIENT. Dear Risk Manager, Datascope / Maquet Getinge is initiating a voluntary Medical Device Recall Removal of the Reinforced Introducer Sets for Maquet 7 Fr., 7.5 Fr.and 8 Fr. IAB (Intra Aortic Balloon) due to a potential breach in the Mylar side (clear plastic) of the pouch, which could compromise device sterility. Identification of the issue: During the execution of a remediation protocol for re-verification of packaging, Datascope / Maquet Getinge found a small slit on the Mylar side (clear plastic) of the pouch of one device. Imprints of the sharp corners of the polystyrene strip used in packaging were observed around the slit, which likely caused the breach in the pouch compromising the sterility of the device. Risk to Health Adequate packaging and packaging materials are essential to help preserve the sterility of medical devices. Improperly sterilized medical devices may pose a serious risk of patient infection, including pyrogenic reaction, inflammatory response, and the possibility of a deadly infection. The probability of serious injury or illness occurring as a result of this potential issue is unlikely, however, likelihood is greater for immunocompromised patients. To date, Datascope / Maquet Getinge has not received any complaints, nor have any adverse events been reported resulting in serious illness or injuries caused by the Reinforced Introducer Set for Maquet 7 Fr., 7.5 Fr.
Initiated
2019-11-20
Root cause
Package design/selection
Terminated
2021-06-15
Distribution
US: AK AL AZ CA CO CT FL GA IA ID IL IN KS KY LA MD MI MN MO MS NC ND NE NJ NY OH OK PA SC SD TN TX VA WA WI WV OUS: Switzerland, Netherlands, South Africa, Canada, Germany, Austria, Spain, Italy, Poland, Bharain, Malaysia
Recall number
Z-0933-2020
Recall status
Terminated
CDRH record ID
178486
FDA record created
2020-02-02
Recall event number
84521

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗