Open original FDA record ↗Recall record
- FDA identifier
- Z-0934-2008
- Company in FDA record
- Inotech Biosystems Intl., Inc
- FDA classification date
- Feb 7, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Erroneous Coagulation Tests
Source facts
- Action
- The recalling firm notified consignees by phone on 08/20/07 and provided instructions for increasing the speed of centrifugation from 1,500 x g to 2,500 x g and decanting the supernatant aliquot carefully so as to not draw in resin fibers that may be stuck on the test tube walls during sample preparation for testing in the coagulation studies. The firm updated their recall notification on 10/01/07 as an Urgent Recall/Correction to alert healthcare practitioners to not use "lupus sensitive reagents" with the heparin absorbant product and that Platelet Factor 3 and any other negatively charged molecule in the plasma may interact with the resin causing false high readings due to the partial removal of the clotting initiating factors.
- Initiated
- 2007-08-20
- Root cause
- Labeling design
- Terminated
- 2012-08-07
- Distribution
- Worldwide; USA to Hospitals and Clinical Laboratories and to one domestic distributor for export to Canada.
- Recall number
- Z-0934-2008
- Recall status
- Terminated
- CDRH record ID
- 65389
- FDA classified
- 2008-02-07
- Recall event number
- 45407