Skip to content

Public FDA record

Z-0934-2008: Inotech brand Heparin Adsorbant, Catalog #: HA-1, in 10mm x 75mm polystyrene tubes, each containing 70mg

Recall record Z-0934-2008 for Inotech Biosystems Intl., Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0934-2008
Company in FDA record
Inotech Biosystems Intl., Inc
FDA classification date
Feb 7, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Erroneous Coagulation Tests

Source facts

Action
The recalling firm notified consignees by phone on 08/20/07 and provided instructions for increasing the speed of centrifugation from 1,500 x g to 2,500 x g and decanting the supernatant aliquot carefully so as to not draw in resin fibers that may be stuck on the test tube walls during sample preparation for testing in the coagulation studies. The firm updated their recall notification on 10/01/07 as an Urgent Recall/Correction to alert healthcare practitioners to not use "lupus sensitive reagents" with the heparin absorbant product and that Platelet Factor 3 and any other negatively charged molecule in the plasma may interact with the resin causing false high readings due to the partial removal of the clotting initiating factors.
Initiated
2007-08-20
Root cause
Labeling design
Terminated
2012-08-07
Distribution
Worldwide; USA to Hospitals and Clinical Laboratories and to one domestic distributor for export to Canada.
Recall number
Z-0934-2008
Recall status
Terminated
CDRH record ID
65389
FDA classified
2008-02-07
Recall event number
45407

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗