An accessory (range shifter, ridge filter&) can be improperly inserted in
snout 300x400 with one rail only and the proton therapy system can still
operate.
Source facts
Action
Ion Beam Applications (IBA) sent an Urgent Field Safety Notice dated December 5, 2017, to customers informing them of the issue and request that all users verify that the accessory is correctly inserted in snout 300x400 with both rails in the guide. A solution to this issue including a hardware modification will be deployed on or before March 31, 2018. For further questions, please call (215) 972-7985.
Initiated
2017-12-05
Root cause
Device Design
Terminated
2019-03-29
Distribution
US Distribution to the states of : MA, FL, VA, PA, OK, TN
Recall number
Z-0934-2018
Recall status
Terminated
CDRH record ID
161711
FDA record created
2018-03-09
Recall event number
79231