Z-0934-2022: Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100

Recall record Z-0934-2022 for Deerfield Imaging, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0934-2022
Company in FDA record
Deerfield Imaging, Inc.
FDA record creation date
Apr 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.

Source facts

Action
The recalling firm issued letters dated 2/17/2022 to its consignees via FedEx and email. The consignee was informed of the issue and that the firm is taking corrective actions to minimize the potential risk of injury, and informed they have been identified as having one or more of the affected mount clamp components that requires replacement. Until the HFD100 mount clamp component is replaced, the firm recommends preventive actions be taken that include: (1) Ensure the instructions in the Operator Manual for proper HFD100 linkage assembly are followed to produce rigid fixation; and (2) Ensure there is no movement of the mount clamp component and of the HFD100 linkages prior to clinical use. The consignee is requested to confirm the details provided on the last page of the letter and return the response to the recalling firm. The details to be confirmed are whether the facility still has the HFD mount clamp component at their facility and requests the serial number and whether it is currently in use.
Initiated
2022-02-17
Root cause
Device Design
Terminated
2024-05-15
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, FL, GA, MA, MN, NC, NY, OH, and TX. The countries of Sweden and Switzerland.
Recall number
Z-0934-2022
Recall status
Terminated
CDRH record ID
192284
FDA record created
2022-04-14
Recall event number
89713

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗