- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Intake filter of the Humidiflow kits may became detached from the module and fall onto the compressor cooling fan during patient treatment. (If the filter were to come in contact with the fan blades this could stop the fan from operating and cause the concentrator to overheat and interrupt the flow of oxygen to the patient)
Source facts
- Action
- Customers were notified via a letter dated April 26, 2007. The letter describes the problem and product. Included with the letter, is a correction kit and installation instructions. The letter also urges the consignee to take an action to fill the attached form to confirm that the new intake filter restaint has been installed.
- Initiated
- 2007-04-26
- Root cause
- Other
- Terminated
- 2011-12-10
- Distribution
- WORLDWIDE DISTRIBUTION: USA including states of GA, MO, FL, UT, CO, TX, MT, MN, PA, OH, AL, SC, VA, KY, IA, MA, TN, IL, NM, UT, NJ, IN, and WI, and countries of CANADA, MEXICO, and JAPAN.
- Recall number
- Z-0935-2007
- Recall status
- Terminated
- CDRH record ID
- 51985
- FDA classified
- 2007-06-20
- Recall event number
- 37863