Open original FDA record ↗Recall record
- FDA identifier
- Z-0935-2014
- Company in FDA record
- Zest Anchors LLC
- FDA classification date
- Feb 6, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend
Source facts
- Action
- Zest Anchors, LLC initiated this recall by sending customers recall notification letter on 01/16/2014 via FedEx overnight. The customer notification letter, titled, "IMPORTANT: MEDICAL DEVICE VOLUNTARY RECALL NOTICE", informed customers of the recall, reason for voluntary recall, product information with part numbers and lot numbers, actions to be taken, and contact information. The customer notification letter was accompanied by a response form titled "MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement & Receipt Form".
- Initiated
- 2014-01-16
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2014-05-15
- Distribution
- Distributed in Ohio and Switzerland.
- Recall number
- Z-0935-2014
- Recall status
- Terminated
- CDRH record ID
- 125247
- FDA classified
- 2014-02-06
- Recall event number
- 67315