Open original FDA record ↗Recall record
- FDA identifier
- Z-0936-2012
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Feb 2, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The skin dose and skin dose rate values are calculated using an incorrect point of reference. Therefore, the AKR as displayed and stored in patient records is too low, although the dose as received by the patient is not affected.
Source facts
- Action
- Philips Healthcare issued on January 9, 2012 a mandatory Field Safety Notice to all affected customers. The letter identified the affected product, problem description, hazard involved and actions that should be taken by the customer/user. The letter contains important information for the continued safe and proper use of the equipment. Customers were instructed to review the following information with all members of there staff and retain a copy with the equipment "Instruction for Use". For information or support concerning this notice contact your local Philips representative 866-767-2822.
- Initiated
- 2012-01-09
- Root cause
- Software design
- Terminated
- 2017-02-13
- Distribution
- Worldwide Distribution - Nationwide Distribution (USA) including the states of AR, AZ, CA, CO, FL, IA, KS, LA, MI, MO, NC, NY, OH, PA, TX, VA, VT, WA and WV
- Recall number
- Z-0936-2012
- Recall status
- Terminated
- CDRH record ID
- 106843
- FDA classified
- 2012-02-02
- Recall event number
- 60936