Open original FDA record ↗Recall record
- FDA identifier
- Z-0938-2013
- Company in FDA record
- Synthes USA HQ, Inc.
- FDA classification date
- Mar 11, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Lot number 7542279, of the 45 mm Click 'X Pedicle Screw (Part number 487.072), was incorrectly etched as "Click 'X Pedicle Screw 05.2 L40 TAN violet" although the actual length of the screw is 45 mm.
Source facts
- Action
- The firm initiated a recall of this product on January 11, 2013 by sending an Urgent Notice: Medical Device Recall letter, dated January 11, 2013, to all consignees. The letter identified the affected product and the issue, as well as requested that consignees examine their inventory for the product(s) with the identified Part and Lot Numbers, remove them from use, and return them to Synthes. Customers were to follow the steps provided in the letter regarding if they had or did not have any of the identified affected product. Questions were to be directed to 610-719-5450 or Fieldaction@synthes.com.
- Initiated
- 2013-01-11
- Root cause
- Employee error
- Terminated
- 2015-09-10
- Distribution
- Distributed in the states of AL, CA, MS, MO, NJ, OR, and TN.
- Recall number
- Z-0938-2013
- Recall status
- Terminated
- CDRH record ID
- 115816
- FDA classified
- 2013-03-11
- Recall event number
- 64173