Skip to content

Public FDA record

Z-0938-2018: Proteus 235, Proton Therapy System a medical device designed to produce and delivers a proton beam for th

Recall record Z-0938-2018 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0938-2018
Company in FDA record
Ion Beam Applications S.A.
FDA record creation date
Mar 9, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a risk of misalignment of the patient when using treatment plans with multiple isocenters.

Source facts

Action
On November 17, 2017 an Urgent Field Safety Notice was issued to customers informing users that when creating or importing a treatment plan with multiple isocenters, the user has to enter Patient Positioning System positions manually.
Initiated
2017-11-17
Root cause
Device Design
Terminated
2021-06-14
Distribution
Nationally
Recall number
Z-0938-2018
Recall status
Terminated
CDRH record ID
161709
FDA record created
2018-03-09
Recall event number
79230

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗