Open original FDA record ↗Recall record
- FDA identifier
- Z-0938-2026
- Company in FDA record
- Instrumentation Laboratory
- FDA classification date
- Dec 17, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Recalled lots were manufactured with double the amount of preservative concentration.
Source facts
- Action
- An URGENT MEDICAL DEVICE RECALL notification, dated 11/18/25, was sent to consignees. Consignees are to contact Client Services at 1-800-955-9525 to arrange conversion to replacement lot N0956748, and in the meantime continue use of affected lots with the QC lot reassignments provided and run QC per labeled instructions for use. Once the replacement lot is received, consignees are to complete the cross-over study per laboratory procedures and immediately destroy affected devices, documenting destruction on the provided Mandatory Urgent Medical Device Recall Tracking Form which should be returned to WERFEN by fax at 781-861-4207 or by email to ra-usa.na@werfen.com. All laboratory staff are to be notified of this recall and a copy of the notice posted near IL Coagulation Systems. Consignees with any questions are to contact Client Services at 1-800-955-9525.
- Initiated
- 2025-11-18
- Root cause
- Process control
- Distribution
- Domestic: CA, CO, FL, NY; Foreign: AR, AT, CA, CZ, DE, DK, ES, FR, GB, IE, IL, IN, IT, JP, LK, NL, PL, PT, RO, RS, TH, VN.
- Recall number
- Z-0938-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217440
- FDA classified
- 2025-12-17
- Recall event number
- 98019