Skip to content

Public FDA record

Z-0938-2026: HemosIL SynthAFax. Partial Thromboplastin Time Tests.

Recall record Z-0938-2026 for Instrumentation Laboratory. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0938-2026
Company in FDA record
Instrumentation Laboratory
FDA classification date
Dec 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Recalled lots were manufactured with double the amount of preservative concentration.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification, dated 11/18/25, was sent to consignees. Consignees are to contact Client Services at 1-800-955-9525 to arrange conversion to replacement lot N0956748, and in the meantime continue use of affected lots with the QC lot reassignments provided and run QC per labeled instructions for use. Once the replacement lot is received, consignees are to complete the cross-over study per laboratory procedures and immediately destroy affected devices, documenting destruction on the provided Mandatory Urgent Medical Device Recall Tracking Form which should be returned to WERFEN by fax at 781-861-4207 or by email to ra-usa.na@werfen.com. All laboratory staff are to be notified of this recall and a copy of the notice posted near IL Coagulation Systems. Consignees with any questions are to contact Client Services at 1-800-955-9525.
Initiated
2025-11-18
Root cause
Process control
Distribution
Domestic: CA, CO, FL, NY; Foreign: AR, AT, CA, CZ, DE, DK, ES, FR, GB, IE, IL, IN, IT, JP, LK, NL, PL, PT, RO, RS, TH, VN.
Recall number
Z-0938-2026
Recall status
Open, Classified
CDRH record ID
217440
FDA classified
2025-12-17
Recall event number
98019

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗