When a previously reported microbiology organism result is deleted on a selected patient, reported susceptibilities will not stay with associated organisms and one or more suceptibilities on the culture will be deleted.
Source facts
Action
Firm sent Product Safety Notice PSN 05-02 on May 10, 2005 which also provides a procedural work-around and an invitation to obtain a service pack to correct the software flaw.
Initiated
2005-05-10
Root cause
Other
Terminated
2005-08-30
Distribution
Nationwide and United Kingdom and Denmark
Recall number
Z-0939-05
Recall status
Terminated
CDRH record ID
39417
FDA classified
2005-06-28
Recall event number
32231
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.