The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
Source facts
Action
Drager sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-800-543-5047 for questions relating to this notice.
Initiated
2012-12-13
Root cause
Labeling design
Terminated
2013-05-20
Distribution
Distributed in the states of CA, IL, and MA.
Recall number
Z-0939-2013
Recall status
Terminated
CDRH record ID
116159
FDA classified
2013-03-11
Recall event number
64447