Public FDA record

Z-0939-2021: FLOW COUPLER Monitor Flow Coupler devices are sold in 5 sizes (2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm).

Recall record Z-0939-2021 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0939-2021
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Jan 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the battery to lose its ability to be recharged.

Source facts

Action
Written notification letter titled, "Urgent Medical Device Recall" was sent to all consignees on 12/28/2020 via first class mail. The letter instructs the consignee to locate all affected devices at their facility; contact Synovis MCA to arrange for product return for the firmware upgrade, power button update, and battery replacement, if applicable; complete the enclosed Baxter customer reply form and return it to Baxter by e-mailing it to fca@baxter.com, even if the consignees does not have any inventory; and if the consignee distributes this product to other facilities or departments within their institution, to please forward a copy of this communication to them.
Initiated
2020-12-28
Root cause
Software design
Distribution
Distributed nationwide to AZ, FL, OH, NY, GA, VA, MA, WI, MO, MT, WV, IL, NC, NJ, AR, CA, CO, CT, IA, ID, IN, KS, LA, MD, MI, MN, MS, NE, NH, NM, NV, OK, OR, PA, RI, WY, WA, VT, TX, TN, SC, SD
Recall number
Z-0939-2021
Recall status
Open, Classified
CDRH record ID
185040
FDA record created
2021-01-30
Recall event number
87013

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗