The reorder number listed on the product label is inaccurate.
Source facts
Action
Consignees were notified by email/fax on 06/17/2005.
Initiated
2005-06-09
Root cause
Other
Terminated
2005-09-21
Distribution
Brazil, Japan & Hong Kong
Recall number
Z-0940-05
Recall status
Terminated
CDRH record ID
40044
FDA classified
2005-06-28
Recall event number
32422
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.