Open original FDA record ↗Recall record
- FDA identifier
- Z-0940-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Feb 2, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During product monitoring firm has discovered that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing necessary. The potential existed for a cable connection to become overheated, leading to the emission of smoke and possible material damage. Firm preventively checked the gradient cabl
Source facts
- Action
- Siemens Medical Solutions sent a "CUSTOMER SAFETY ADVISORY NOTICE" letter dated November 24, 2010 to all affected customers. The letter described the product, problem, and actions to be taken. Contact the firm at 610-219-4834 for questions regarding this recall.
- Initiated
- 2012-01-26
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-01-31
- Distribution
- Nationwide Distribution-including the states of AK, AZ, CA, CT, DE, FL, GA, IL, IN, KS, MD, MA, MI, MN, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, and WI.
- Recall number
- Z-0940-2012
- Recall status
- Terminated
- CDRH record ID
- 107139
- FDA classified
- 2012-02-02
- Recall event number
- 61030