Skip to content

Public FDA record

Z-0940-2017: TVC Imaging System model TVC-MC9 , for the near-infrared examination of coronary arteries in patients und

Recall record Z-0940-2017 for Infraredx, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0940-2017
Company in FDA record
Infraredx, Inc.
FDA record creation date
Jan 3, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demonstrated an unusual appearance in that the yellow in the image (indicating the presence of lipid core plaques) was horizontally-oriented

Source facts

Action
Infraredx, Inc. sent an Urgent Product Recall Medical Device Field Correction letter dated December 15, 2016, describing the problem and replacing the defective lasers. A Field Service Representative will contact customers to schedule a replacement of the system's laser rack module. If customers receive this letter and are not contacted within 48 hours, please call Infraredx Technical Support@ 1-800-596-3104. For questions regarding this recall call 781-221-0053.
Initiated
2016-12-15
Root cause
Device Design
Terminated
2018-10-09
Distribution
Worldwide Distribution - US including MI and Internationally to South Korea
Recall number
Z-0940-2017
Recall status
Terminated
CDRH record ID
151938
FDA record created
2017-01-03
Recall event number
76014

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗