Reason stated in the source
Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demonstrated an unusual appearance in that the yellow in the image (indicating the presence of lipid core plaques) was horizontally-oriented
Source facts
- Action
- Infraredx, Inc. sent an Urgent Product Recall Medical Device Field Correction letter dated December 15, 2016, describing the problem and replacing the defective lasers. A Field Service Representative will contact customers to schedule a replacement of the system's laser rack module. If customers receive this letter and are not contacted within 48 hours, please call Infraredx Technical Support@ 1-800-596-3104. For questions regarding this recall call 781-221-0053.
- Initiated
- 2016-12-15
- Root cause
- Device Design
- Terminated
- 2018-10-09
- Distribution
- Worldwide Distribution - US including MI and Internationally to South Korea