Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.
Source facts
Action
On 01/05/2021 Biomeme, Inc., issued an Urgent Device Recall Correction for the latest build of the Franklin Real-Time PCR Thermocycler, which affects some devices built since September 21st, 2020. This software correction is to fix an issue where the device is unable to get past the beginning stages of the test.
Initiated
2021-01-05
Root cause
Software design
Terminated
2021-06-11
Distribution
Domestic: FL, TX, IN, KY, OH, NY, CA, PA, MD. Foreign: Canada, Zimbabwe.
Recall number
Z-0940-2021
Recall status
Terminated
CDRH record ID
185447
FDA record created
2021-02-01
Recall event number
87153