Z-0941-2009: Stryker InTouch 1.0 Zoom Critical Care Bed; Model 2140. The device is intended to be used as an Intensiv

Recall record Z-0941-2009 for Stryker Medical Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0941-2009
Company in FDA record
Stryker Medical Div. of Stryker Corporation
FDA classification date
Feb 25, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The touch screen on the footboard may display a "tilt over range" error, resulting in a precautionary stop of the bed motions which result in the bed appearing to have locked out some of its bed positioning features.

Source facts

Action
Consignees were notified of the problems via letter dated January 5, 2009, advised to check their beds and to take them out of service if the foot board touchscreen displayed a "tilt over range" error, and that a Stryker Medical field service representative would contact them to perform software upgrades on the affected beds. Please direct questions to Stryker Medical at 1-269-324-6644.
Initiated
2009-01-05
Root cause
Device Design
Terminated
2010-11-05
Distribution
Worldwide Distribution including USA and countries of Canada, Japan and Switzerland.
Recall number
Z-0941-2009
Recall status
Terminated
CDRH record ID
74681
FDA classified
2009-02-25
Recall event number
49610

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗