Open original FDA record ↗Recall record
- FDA identifier
- Z-0941-2012
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Feb 2, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
It has been found that some of the starter cables used for x-ray tube high speed rotation are not sufficiently durable.
Source facts
- Action
- The recall notification titled "URGENT: MEDICAL DEVICE CORRECTION", was sent on 12/29/11 to customers via certified mail. The letter informed the customer of an issue with the starter cable used for high speed rotation on the Radrex X-Ray System. The letter also informs customers of the corrective measures: a replacement cable will be installed on your system to prevent this issue from occurring. An 800 number was provided for any questions regarding the letter.
- Initiated
- 2011-12-29
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-03-24
- Distribution
- Class II Recall - Nationwide Distribution (USA) - Including the states of VA, IL, CA, PA, TN, OH, LA, TX, OK, AR, AZ, MS, WV, TN, WI, NJ, MN, NY, IA, FL, and Washington D.C.
- Recall number
- Z-0941-2012
- Recall status
- Terminated
- CDRH record ID
- 107036
- FDA classified
- 2012-02-02
- Recall event number
- 61010