Recall record Z-0941-2026 for DICOM Grid, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0941-2026
Company in FDA record
DICOM Grid, Inc.
FDA classification date
Dec 18, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar reconstruction applied; Manual rotation applied, could result in inaccurate length measurement tool values that could compromise diagnostic accuracy, lead to misdiagnosis or inappropriate follow-up.

Source facts

Action
On 11/18/2025, correction notices were emailed to customers informing them of the following: 1) A stop deploy was issued for the affected versions. A software patch release will remove the use of measurements within the ProViewer when the user has both applied the MPR view and made a rotation on the plane. User manuals and guides will be updated to remove reference to this feature. 2) Until the update is applied, advise all staff to refrain from using measurements in MPR mode after rotating one of the planes. 3) Complete and return the acknowledgement form via email to email to regulatory-affairs@intelerad.com A software patch release to address this issue will be applied to all affected InteleShare servers between Nov 18-20th, 2025. There will be no service disruption associated with this release. For more information or to ask any questions about this notification, please contact the firm's regulatory affairs department at regulatory-affairs@intelerad.com
Initiated
2025-11-18
Root cause
Software design
Distribution
US: AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ, VA, WA, NC, TN, WI, GA, AK, CA, SD, CO, MA, WV, MN, KS, MD, VT, IA, RI, AL, ID, LA, PR, NH, IN, OK, UT, NE, SC, AR, DE OUS: CA, NZ, AU, GB, BE, FR, IT, NO, SE
Recall number
Z-0941-2026
Recall status
Open, Classified
CDRH record ID
217401
FDA classified
2025-12-18
Recall event number
97993

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗