The touch screen on the footboard may display a "tilt over range" error, resulting in a precautionary stop of the bed motions which result in the bed appearing to have locked out some of its bed positioning features. Also, the "hold" button, used to interrupt an intended patient turn on beds with an integrated PositionPro mattress, may only function intermittently, resulting in an inability to ho
Source facts
Action
Consignees were notified of the problems via letter dated January 5, 2009, advised to check their beds and to take them out of service if the foot board touchscreen displayed a "tilt over range" error, and that a Stryker Medical field service representative would contact them to perform software upgrades on the affected beds. Please direct questions to Stryker Medical at 1-269-324-6644.
Initiated
2009-01-05
Root cause
Device Design
Terminated
2010-11-05
Distribution
Worldwide Distribution including USA and countries of Canada, Japan and Switzerland.
Recall number
Z-0942-2009
Recall status
Terminated
CDRH record ID
74682
FDA classified
2009-02-25
Recall event number
49610