Open original FDA record ↗Recall record
- FDA identifier
- Z-0942-2014
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Feb 9, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ortho Clinical Diagnostics issued an Urgent Product Correction Notification involving the enGen Select V5.0 & enGen Custom v3.2.2 due to an anomaly that affects the autoverification feature and custom rules that depend upon data elements used by the referenced enGen Systems.
Source facts
- Action
- Ortho Clinical Diagnostics (OCD), issued an Urgent Product Correction Notification letter dated October 7, 2013 and on December 17, 2013 to their customers via Federal Express or via US Postal Service. The letter identified the affected product, problem and actions. The notices advise the customers to ensure that all required data elements are entered and in the correct format if using the autoverification or custom rules for the referenced enGen Laboratory Automation Systems. OCD trained service representatives will contact all affected customers. Customers with questions may call the Customer Technical Service at 1-800-421-3311.
- Initiated
- 2013-10-08
- Root cause
- Software design
- Terminated
- 2017-04-11
- Distribution
- Worldwide Distribution - USA Nationwide in the states and countries of: AZ, FL, WV, TX, PA, DE, MI, OH, TN, IL, NC, Chile, India, Italy, Portugal, Spain, Sweden, UK
- Recall number
- Z-0942-2014
- Recall status
- Terminated
- CDRH record ID
- 122442
- FDA classified
- 2014-02-09
- Recall event number
- 66467