Open original FDA record ↗Recall record
- FDA identifier
- Z-0942-2017
- Company in FDA record
- Pride Mobility Products Corp
- FDA classification date
- Jan 3, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The interface between the TB3 back and tilt base can separate when tilted beyond the specified angle.
Source facts
- Action
- Pride Mobility Products Corp. mailed a letter to affected customers to inform them of the issue. Systems were not required to be returned. Customers were informed that a TB3 Seat Depth Adjustment Reinformcement Kit which contains (2) reinforcement plates and associated hardware to strengthen the interface will be installed to address the issue. Direct accounts were asked to contact patients to inform them of the product correction. Customers were also requested to complete and return Pride's completion card once the correction has been made. Pride will issue a $65.00 credit once the completion card is received. Customers with questions were instructed to call Pride's Upgrade Department at 866-222-6212. For questions regarding this recall call 800-800-8586.
- Initiated
- 2016-12-01
- Root cause
- Use error
- Terminated
- 2020-03-19
- Distribution
- Nationwide Distribution
- Recall number
- Z-0942-2017
- Recall status
- Terminated
- CDRH record ID
- 151534
- FDA classified
- 2017-01-03
- Recall event number
- 75881