Open original FDA record ↗Recall record
- FDA identifier
- Z-0942-2018
- Company in FDA record
- Draegar Medical Systems, Inc.
- FDA record creation date
- Mar 9, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Drger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.
Source facts
- Action
- In September of 2017 an Urgent medical Device Recall letter was issued to customers informing them of the recall and to request that inventory be checked to see if the product subject in the recall is on the premises. To do this, the customer will check the label for the revision index number which is located on the upper middle portion of label. RI number 15 and 16 are included in the recall. No other numbers are subject to recall. Affected product will receive a sleeve from the company to correct the issue. Questions or concerns can be directed to 1-800-437-2437 (Press 1, then press 32349) or contact Customer Care at 1-800-437-2437 (Press 4, then press 4 again)
- Initiated
- 2017-10-04
- Root cause
- Nonconforming Material/Component
- Terminated
- 2018-06-15
- Distribution
- Nationally
- Recall number
- Z-0942-2018
- Recall status
- Terminated
- CDRH record ID
- 161659
- FDA record created
- 2018-03-09
- Recall event number
- 79209