There is an unsecure connection between the male Luer and the stopcock on the I.V. extension set.
Source facts
Action
Recall letters dated 4/21/05 were sent to the direct accounts via first class mail on the same date to the attention of the Director of Materials Management. The accounts were informed of the reports of an unsecure connection between the male luer and the stopcock due to silicone on the male luer tip, and were requested to examine their inventory for the affected lots of product, and call Baxter Healthcare Center for Service at 1-888-229-001 to return the affected product for credit.
Initiated
2005-04-21
Root cause
Other
Terminated
2006-07-19
Distribution
Nationwide and internationally to American Samoa and Venezuela.
Recall number
Z-0943-05
Recall status
Terminated
CDRH record ID
38524
FDA classified
2005-06-30
Recall event number
31800
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.