Skip to content

Public FDA record

Z-0943-2018: Proteus 235, Pencil Beam Scanning, PTS-9.4.X, PTS-10.0.1.1 The Proton Therapy System - Proteus 235 (bran

Recall record Z-0943-2018 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0943-2018
Company in FDA record
Ion Beam Applications S.A.
FDA record creation date
Mar 9, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The internal configuration of the electrometers and the Real-Time control boards of the Dekimo Scanning Controller is not checked before each patient treatment.

Source facts

Action
The firm, IBA, sent an "Urgent Field Safety Notice" dated 10/31/2017 to customers on 10/31/2017. The notice described the product, problem and actions to be taken. stating the following: There is no specific recommendation for the user. IBA will provide an Internal User notice to the IBA site team, requesting the operator to select the Beam Delivery Point CGTR, by clicking on the BDP button, before each patient treatment preparation (see Annex 1 for details). This will begin the check of the internal configuration of all the elements of the Dekimo Scanning Controller. An updated system will be deployed on February 28, 2017. If you have any questions, contact Customer Complaints & Vigilance Director at Vigilance@iba-group.com or +32 10 203 787 or helpdesk at +32 2 507 20 81 (available 24/7).
Initiated
2017-11-01
Root cause
Device Design
Terminated
2019-03-29
Distribution
Worldwide Distribution: US (nationwide) to: MI only and country of: France.
Recall number
Z-0943-2018
Recall status
Terminated
CDRH record ID
161706
FDA record created
2018-03-09
Recall event number
79229

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗