- FDA identifier
- Z-0943-2018
- Company in FDA record
- Ion Beam Applications S.A.
- FDA record creation date
- Mar 9, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The internal configuration of the electrometers and the Real-Time control boards of the Dekimo Scanning Controller is not checked before each patient treatment.
Source facts
- Action
- The firm, IBA, sent an "Urgent Field Safety Notice" dated 10/31/2017 to customers on 10/31/2017. The notice described the product, problem and actions to be taken. stating the following: There is no specific recommendation for the user. IBA will provide an Internal User notice to the IBA site team, requesting the operator to select the Beam Delivery Point CGTR, by clicking on the BDP button, before each patient treatment preparation (see Annex 1 for details). This will begin the check of the internal configuration of all the elements of the Dekimo Scanning Controller. An updated system will be deployed on February 28, 2017. If you have any questions, contact Customer Complaints & Vigilance Director at Vigilance@iba-group.com or +32 10 203 787 or helpdesk at +32 2 507 20 81 (available 24/7).
- Initiated
- 2017-11-01
- Root cause
- Device Design
- Terminated
- 2019-03-29
- Distribution
- Worldwide Distribution: US (nationwide) to: MI only and country of: France.
- Recall number
- Z-0943-2018
- Recall status
- Terminated
- CDRH record ID
- 161706
- FDA record created
- 2018-03-09
- Recall event number
- 79229