Open original FDA record ↗Recall record
- FDA identifier
- Z-0943-2021
- Company in FDA record
- The Binding Site Group, Ltd.
- FDA record creation date
- Feb 1, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to assigned values.
Source facts
- Action
- On 11/30/2020, the firm emailed an "Urgent Field Safety Notice" to its U.S. Customers informing them that investigations have confirmed that the quality control (QC), high and low controls, provided with specific product and lot numbers are consistently demonstrating a ~15% positive bias compared to assigned values and customers may experience QCs consistently reporting results close to the upper acceptable limits, and QC failures may occur on the affected products. Customer are instructed: 1. Users may continue to use any remaining kits of the affected product in their stock. As per the product package insert sample results obtained should only be accepted if the control results are within +/-20% of the concentrations stated on the QC certificate provided. 2. To complete the E-Back Form TSWS18 to Technical.support@bindingsite.com or to their local representative within one week of receiving this notification. Customers are also instructed: -To pass on the notification to all those who need to be aware within their organization or to any organization where the potentially affected devices have been transferred. -To transfer this notification to other organizations on which the information has an impact. -To maintain awareness on this notification for any appropriate period. Any questions, contact the local representative or Technical Support Group: UK technical support is: Telephone: +44(0) 1214569696 Email: technical.support@bindingsite.com Customers in the USA may choose to instead contact the local representative at: Telephone: 800-633-4484 Us-tsc@Bindingsite.com
- Initiated
- 2020-11-18
- Root cause
- Under Investigation by firm
- Terminated
- 2024-12-23
- Distribution
- U.S.: CO, FL, IA, MA, NC, NJ, NY, OH, SD, and UT. O.U.S.: Not provided
- Recall number
- Z-0943-2021
- Recall status
- Terminated
- CDRH record ID
- 184840
- FDA record created
- 2021-02-01
- Recall event number
- 86926