Skip to content

Public FDA record

Z-0943-2021: Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A

Recall record Z-0943-2021 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0943-2021
Company in FDA record
The Binding Site Group, Ltd.
FDA record creation date
Feb 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to assigned values.

Source facts

Action
On 11/30/2020, the firm emailed an "Urgent Field Safety Notice" to its U.S. Customers informing them that investigations have confirmed that the quality control (QC), high and low controls, provided with specific product and lot numbers are consistently demonstrating a ~15% positive bias compared to assigned values and customers may experience QCs consistently reporting results close to the upper acceptable limits, and QC failures may occur on the affected products. Customer are instructed: 1. Users may continue to use any remaining kits of the affected product in their stock. As per the product package insert sample results obtained should only be accepted if the control results are within +/-20% of the concentrations stated on the QC certificate provided. 2. To complete the E-Back Form TSWS18 to Technical.support@bindingsite.com or to their local representative within one week of receiving this notification. Customers are also instructed: -To pass on the notification to all those who need to be aware within their organization or to any organization where the potentially affected devices have been transferred. -To transfer this notification to other organizations on which the information has an impact. -To maintain awareness on this notification for any appropriate period. Any questions, contact the local representative or Technical Support Group: UK technical support is: Telephone: +44(0) 1214569696 Email: technical.support@bindingsite.com Customers in the USA may choose to instead contact the local representative at: Telephone: 800-633-4484 Us-tsc@Bindingsite.com
Initiated
2020-11-18
Root cause
Under Investigation by firm
Terminated
2024-12-23
Distribution
U.S.: CO, FL, IA, MA, NC, NJ, NY, OH, SD, and UT. O.U.S.: Not provided
Recall number
Z-0943-2021
Recall status
Terminated
CDRH record ID
184840
FDA record created
2021-02-01
Recall event number
86926

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗