Open original FDA record ↗Recall record
- FDA identifier
- Z-0944-2011
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Feb 7, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Although the system x-ray output is accurate for the technique factors given, and the dose display is also accurate, certain CTDI adjustment factors used for the manual calculation from the Technical Reference Manual (TRM) are incorrect. CTDI100 Aperture Adjustment Factors for 4x0.625 mode contain incorrect values. This issue may lead to inaccurate CTDI manual calculation. The results of the diffe
Source facts
- Action
- An Important Electronic Product Radiation Warning letter will be provided together with the TRM addendum informing the customers of the issue and related hazards. The letter also identified the affected product and gave instructions, which include referring to the TRM addendum, along with the product correction. This correction will be executed via a Field Modification Instruction IFMI 254281 released in November, 2010 with an estimated completion date of May, 2011. Questions or concerns should be directed to the call numbers listed or a local GE Healthcare Service Representative for countries not listed.
- Initiated
- 2010-10-27
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2012-07-13
- Distribution
- Worldwide Distribution
- Recall number
- Z-0944-2011
- Recall status
- Terminated
- CDRH record ID
- 95567
- FDA classified
- 2011-02-07
- Recall event number
- 57193