Z-0944-2021: Philips EPIQ Diagnostic Ultrasound System; Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C,

Recall record Z-0944-2021 for Philips Ultrasound Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0944-2021
Company in FDA record
Philips Ultrasound Inc
FDA record creation date
Feb 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The manufacturer discovered an issue associated with the EPIQ Image Boost with xPlane and Color Flow or Doppler while using the X8-2t TEE Transducer. If Image Boost is enabled the color box shows flow visualization not accurately represent the fluid flow. Similarly, if Image Boost is enabled, both CW and PW Doppler traces will not accurately represent the fluid flow.

Source facts

Action
On 12/16/2020, the manufacturer sent an Urgent Medical Device Correction letter (MDC) with instructions on how to turn off the Image Boost feature. Philips will resolve the issue by providing a software update, at no cost, to correct all affected EPIQ Ultrasound Systems (the update will disable the Image Boost feature when connected to an X8-2t TEE Transducer). Method of Notification - Letter (Urgent Medical Device Correction Letter in the US, Field Safety Notice in rest of the world, Distributor Letter in rest of the world). For US customers, the letters will be sent via certify mail with electronic return receipt. Philips Markets Organizations are responsible for distributing the letters outside of the US.
Initiated
2020-12-16
Root cause
Under Investigation by firm
Terminated
2023-03-15
Distribution
US - AL, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE,NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI and WV OUS- Austria, Australia, Belgium, Canada, Chezh Republic, Croatia, Denmark, Egypt, Finland, France, Hong Kong, Indonesia, India, Ireland, Italy, Japan, Latvia, Lithuania, Malta, Mexico, Netherlands, Norway, Poland, Romania, Russia, Senegal, Slovenia, Sri Lanka, South Africa, Switzerland, Thailand, Turkey, UAE and United Kingdom.
Recall number
Z-0944-2021
Recall status
Terminated
CDRH record ID
185039
FDA record created
2021-02-02
Recall event number
87012

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗