Open original FDA record ↗Recall record
- FDA identifier
- Z-0945-2008
- Company in FDA record
- Boston Scientific
- FDA classification date
- Jan 31, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Flash may be protruding from the lumen of the catheter shaft. If this flash is present in the catheter, there is the potential of an embolization resulting from the flash detaching from the device during the procedure.
Source facts
- Action
- The firm sent an Urgent Medical Device Recall letter dated 11/20/2007, to all affected hospitals. The letter describes the defect with the catheters and Product Table information including UPNs, Catalog Numbers, Lot/Batch Numbers and Product Descriptions. The firm requests that further distribution or use of any remaining product affected by the recall should cease immediately. ****IMPORTANT INFORMATION: If it is a practice of the hospital to remove product from the outer carton and store on shelves in the inner-pouch only. Users are to review the Product Table carefully and consider both the inner and outer packaging product codes when search for affected/recalled product.****
- Initiated
- 2007-11-20
- Root cause
- Release of Material/Component prior to receiving test results
- Terminated
- 2008-10-23
- Distribution
- Nationwide Distribution including states of AL, CA, FL, GA, KS, KY, LA, MN, NV, OH, OK, PA, SD, TN, TX, and WI.
- Recall number
- Z-0945-2008
- Recall status
- Terminated
- CDRH record ID
- 68033
- FDA classified
- 2008-01-31
- Recall event number
- 45903