Skip to content

Public FDA record

Z-0945-2021: Tourni-Cot - Universal, UDI 00855364004085, Model Number TCU, Non-pneumatic Tourniquet - Product Usage:

Recall record Z-0945-2021 for Mar-Med Co. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0945-2021
Company in FDA record
Mar-Med Co
FDA classification date
Feb 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A failure can occur if the user attempts to remove the device by partially snipping the edge and pulling on it. The device may tear into two pieces with a portion remaining in place, which can continue to constrict the vessels. This type of failure can result in digit ischemia.

Source facts

Action
On December 28, 2020, the firm began notifying customers of the recall via Urgent Medical Device Correction letters. Customers were informed that serious injury could occur due to the failure mode associated with the correction. Actions to be taken by the Customer: Ensure proper removal of the Uni-Cot device by completely severing it from the digit, followed by verification that none of the device is remaining. Actions being taken by Mar-Med: Phase 1: Include a notice in each box of the device that identifies the risk for the users of this type of failure and provides specific guidance on how to avoid it. Phase 2: Include a warning tag on the device. The warning tag will not release unless the device is completely removed from the digit. Phase 3: In the near future an update will be made to the design of the device to better address usability and removal, maintaining the inclusion of the warning tag. An updated IFU will also be designed to provide clearer instructions and warnings. Please contact Mar-Med with any questions or information about this type of failure at 616-454-3000 or at info@marmed.com.
Initiated
2020-12-28
Root cause
Device Design
Terminated
2023-09-14
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Canada, France, Germany, Greece, Italy, Japan, Netherlands, Singapore, Spain, Sweden, Switzerland, and United Kingdom.
Recall number
Z-0945-2021
Recall status
Terminated
CDRH record ID
185036
FDA classified
2021-02-02
Recall event number
87009

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗