Z-0946-2013: Nicolet EEG Wireless Amplifier 32/64 Channel. Catalog/Part Numbers : 515-015400 for the 32 channel amplif

Recall record Z-0946-2013 for Natus Neurology Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0946-2013
Company in FDA record
Natus Neurology Incorporated
FDA classification date
Mar 13, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Natus Neurology Incorporated is recalling the Nicolet EEG Wireless Amplifier which may overheat and become uncomfortable to touch around the area of the on/off button. The overheating may result in discomfort ranging from too hot up to the potential for a burn.

Source facts

Action
Natus sent an "Urgent Medical Device Correction and Removal Notification' letter dated February 27, 2013 to all affected customers. The letter described the affected product, problem and actions to be taken. The letter advised customers that a Natus Field Service technician will remove and replace the affected product at no charge to the users. The letter also advised if any of the affected products have been transferred to another location to please forward a copy of the letter and to notify Natus. For questions call 800-356-0007 ext. 5129.
Initiated
2013-02-27
Root cause
Nonconforming Material/Component
Terminated
2014-04-25
Distribution
Worldwide Distribution - USA Nationwide including the states of CA, FL, LA, MD, MS, MO, NJ, NY, PA, TX and the country of AUSTRALIA. BAHRAIN BRAZIL CANADA Country DENMARK EGYPT GERMANY HONG KONG INDIA INDONESIA IRELAND ITALY JAPAN KOREA KUWAIT MALAYSIA MEXICO NETHERLANDS ROMANIA RUSSIA SINGAPORE SPAIN SWITZERLAND TAIWAN THAILAND TURKEY UNITED KINGDOM URUGUAY VIET NAM
Recall number
Z-0946-2013
Recall status
Terminated
CDRH record ID
116224
FDA classified
2013-03-13
Recall event number
64479

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗